Cluster check on fluorescent microscope
Prepare reagents
in advance thaw 2ml aliquote of staining buffer (SB);
combine 2ml SB + 0.4µl SYBR Green I and mix;
label and fill in strips:
NB! do not prepare #S strips, if the FC is stained directly after amplification; storage buffer is in position 12
2x #S : 8x 160µl of storage buffer,
2x #Y : 8x 120µl of staining solution (for the control FC)
#W : 8x 240µl of H2 O,
#O : empty
check you have:
staining manifold
control FC with clusters
2x Petri dishes
small pieces of parafilm
Flowcell staining
install staining manifold:
after amplification
air gap [26 · 40µl/min · 120µl] & remove amplification manifold;
install staining manifold;
after storage
turn on CS;
wait 30s, turn on computer;
check/empty waste container;
air gap [26 · 40µl/min · 120µl] & remove washing bridge;
install FC & staining manifold;
install tube strip holder;
staining:
strip #Y (staining solution) [26 · 30µl/min · 100µl];
strip #0 (air gap), [26 · 40µl/min · 25µl];
unclip "reagent delivery port" & "central clamp" & rise up staining manifold;
remove FC;
install control FC & attach the staining manifold back;
repeat step 3 for control FC;
install empty FC and attach the staining manifold back;
prepare FCs for imaging:
clean FCs with water and lens cleaning tissue;
put FCs in Petri dishes, close holes with pieces of parafilm;
take into microscope room:
FCs;
lens cleaning tissue;
wash bottle;
Fluorescent microscope
take pictures from all channels:
NB! UV is bad for clusters, be quick
open shutter;
adjust the focus to see bottom clusters;
take the picture: "Snap" in the "Live" window;
close shutter;
save picture in the FC folder (format: solexa_2008_10_16_chX.tif);
change Y position in the MW to move to another channel;
finish:
click "Load position" on the touchscreen (Fig.2: 5);
put the test FC back in Petri dish;
close the AxioVision program;
transfer files to your computer;
if you are the last user:
turn off the Power Supply (Fig.1: 3);
turn off the FluoArc lamp (Fig.1: 2);
logout.
Flowcell wash for storage
fill stained FCs with storage solution:
after storage
for each stained FC:
unclip "reagent delivery port" and "central clamp" of staining manifold;
remove empty FC;
install the stained FC & attach the staining manifold back;
strip #S(storage solution) [26 · 40µl/min · 120µl];
strip #0(air) [26 · 40µl/min · 25 µl];
unclip "reagent delivery port" and "central clamp" of staining manifold;
remove FC
after amplification
clean FC with water and lens cleaning tissue;
put flowcell in 50ml tube with storage buffer, store at 4°C;
wash empty FC and staining manifold:
install empty FC & attach the staining manifold back;
strip #W(H2 O) [26 · 60µl/min · 240µl];
strip #0(air) [26 · 40µl/min · 120µl];
remove empty FC & staining manifold;
install bridge manifold;
Flowcell processing for sequencing
wash empty FC and staining manifold:
strip #W(H2 O) [26 · 60µl/min · 240µl];
strip #0(air) [26 · 40µl/min · 120µl];
remove empty FC & staining manifold;
clean flowcell area with H2 O and lens cleaning tissues;
install "linearization/blocking/hybridization" reagents;
fill control FC with storage solution:
install control FC & attach the staining manifold back;
storage solution [12 · 40µl/min · 120µl];
air gap [26 · 40µl/min · 25 µl];
unclip "reagent delivery port" and "central clamp" of staining manifold;
remove control FC clean it with water and lens cleaning tissue;
put control FC in 50ml tube with storage buffer, store at 4°C;
install test FC;
storage solution [12 · 40µl/min · 120µl];
go to linearization/blocking/hybridization protocol;
Solutions
SYBR Green I
heat stock solution to NT;
prepare 0.4µl aliquotes in 0.5ml tubes;
freeze in liquid nitrogen and store at -20°C dessicated;
Notes
Fluorescent microscope IMAGER.Z1 (Zeiss Axioimager.Z1) T1, Room 22
online booking : date and time
computer : Lab-RR/Mg04-33
responsible : Rudi Lurz; T1, R308; phone (+49 30) 8413 1644; lurz@molgen.mpg.de
keys : abt. Herrmann, Rudi Lurz, Bodo Lange, Sylvia Krobitsch
Fig. 1. 1 — lamp regulator; 2 — UV-lamp; 3 — microscope power supply; 4 — optic cleaning pads; 5 — optic cleaning sticks; 6 — image output selector; 7 — microscope table.
Fig. 2. 1 — microscope table; 2 — image output selector; 3 — XY-ajustment; 4 — Z-ajustment; 5 — touchscreen.
Fig. 3. FC-position in the right corner of the slide holder.
Second-generation sequencing
URL:
http://seq.zbio.net
e-mail:
soldatov@molgen.mpg.de
visits:
Warning : require(/home/molbiol/data/www/vphp/include.php) [
function.require ]: failed to open stream: No such file or directory in
/usr/home/molbiol/domains/molbiol.ru/public_html/seq/ssi/counter.php on line
6
Fatal error : require() [
function.require ]: Failed opening required '/home/molbiol/data/www/vphp/include.php' (include_path='.:/usr/local/lib/php') in
/usr/home/molbiol/domains/molbiol.ru/public_html/seq/ssi/counter.php on line
6